4.3 Article

Colonic anastomotic stenoses and Memotherm stent fracture: A report of three cases

期刊

CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY
卷 24, 期 5, 页码 336-339

出版社

SPRINGER-VERLAG
DOI: 10.1007/s00270-001-0033-7

关键词

colon, stenosis and obstruction; stents and prostheses; interventional procedures

向作者/读者索取更多资源

Deployment of a Memotherm colonic stent (Bard, Angiomed, Karlsruhe, Germany) across anastomotic strictures, following anterior resection, is described in three patients. Two patients presented with symptoms of colonic obstruction. Two of the patients had previously undergone. unsuccessful balloon dilatation of the stricture. In the third, in addition to the anastomotic stricture, there was local tumor recurrence. Initially, stenting provided effective relief of symptoms. However, in all three patients, fracture of the stents occurred at intervals of 3-7 months after insertion. This use and complication of colonic stenting has not been reported previously.