Metallic neurosurgical implants: Evaluation of magnetic field interactions, heating, and artifacts at 1.5-tesla

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 14, 期 3, 页码 295-299

出版社

WILEY
DOI: 10.1002/jmri.1185

关键词

magnetic resonance imaging; safety; magnetic resonance imaging, implants; heating; artifacts

向作者/读者索取更多资源

The purpose of this study was to use ea vivo testing to determine the magnetic resonance imaging (MRI) safety aspects for seven different metallic neurosurgical implants in association with the 1.5-T MR environment. Ex vivo testing was performed using previously-described techniques for the evaluation of magnetic field interactions (deflection angle and torque), heating (gel-filled phantom and fluoroptic thermometry; 15 minutes of MRI at a specific absorption rate [SAR] of 1.4 W/kg), and artifacts (using T1-weighted, spin-echo and gradient-echo pulse sequences). None of the metallic implants displayed interactions with the magnetic field. The highest temperature change was +0.6degreesC for the representative implant that was evacuated. Artifacts were relatively minor. The lack of magnetic field interactions and negligible heating indicate that MR procedures may be conducted safely in patients with these neurosurgical implants using MR systems with static magnetic fields of 1.5-T or less. Further-more, these implants may be considered for use in interventional MR procedures insofar as the MR safe qualities and relatively small artifacts would likely be desirable for such procedures. J. Magn. Reson. Imaging 2001;14: 295-299. (C) 2001 Wiley-Liss, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据