Enhancing the specificity of the COBAS AMPLICOR CT/NG test for Neisseria gonorrhoeae by retesting specimens with equivocal results

期刊

JOURNAL OF CLINICAL MICROBIOLOGY
卷 39, 期 9, 页码 3092-3098

出版社

AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.39.9.3092-3098.2001

关键词

-

向作者/读者索取更多资源

The COBAS AMPLICOR CT/NG test for Neisseria gonorrhoeae cross-reacts with certain strains of nonpathogenic Neisseria species. In some strains, the target sequence is identical to that of N. gonorrhoeae, whereas other strains have a small number of mismatches within the regions recognized by the primers or probe used in the COBAS AMPLICOR NG test. These cross-reactive strains are occasionally present in urogenital specimens, causing false-positive results in the COBAS AMPLICOR NG test. Analysis of the data generated in a large multicenter clinical trial showed that 2.9% of the specimens gave signals between A(660)s of 0.2 and 3.5 but that one-half of these equivocal specimens did not contain N. gonorrhoeae. Most of these equivocal specimens were correctly classified as true positive or true negative by retesting in duplicate and defining a PCR-positive result as two of three results with an A(660) of greater than or equal to2.0. If specimens had been classified as positive or negative based on a single test result using a cutoff of an A(660) of 0.2, specificity would have ranged from 96.2 to 98.9% depending on specimen type, sex, and presence of symptoms. By employing the equivocal zone-retesting algorithm, specificity increased to 98.6 to 99.9% with little effect (0.1 to 4.9% decrease) on sensitivity in most specimen types, enabling the test to achieve a positive predictive value of at least 90% in populations with a prevalence of 4% or higher. In lower-prevalence populations, the test could be used to screen for presumptive infections that would have to be confirmed by an independent test.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据