4.7 Article

Prussian Blue bulk modified screen-printed electrodes for H2O2 detection and for biosensors

期刊

TALANTA
卷 55, 期 3, 页码 605-611

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0039-9140(01)00469-6

关键词

Prussian blue; screen-printed electrode; hydrogen peroxide; biosensors; glucose oxidase; mediator

向作者/读者索取更多资源

A sensor for H2O2 amperometric detection based on a Prussian Blue (PB) bulk modified carbon screen-printed electrode was developed. It has been optimised with respect to the lowest limit of detection achieved. PB was made chemically by the reaction of FeCl3 with K-4[Fe(CN)(6)]. The resulting powder, obtained by forced crystallisation induced by acetone, was dried and activated at 150 degreesC for 10 h. PB microparticles (< 38 mum) were prepared and mixed with carbon ink. The limit of detection achieved was 0.4 muM with the linear range up to 100 muM of H2O2 with the sensitivity of 137 muA mM(-1) cm(-1), that was comparable with sensors based on electrodeposited PB film. The transducer was applied for a glucose biosensor, that exhibited LOD of 0.22 mM, linear range up to 3 mM, K, K-M(app) of 4.6 mM, and the sensitivity of 3.21 +/- 0.16 muA mM(-1) cm(-1). The peroxide sensor, as well as the glucose biosensor, were totally insensitive to oxygen, ascorbate, urate, and paracetamol. (C) 2001 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据