4.7 Article

Prussian Blue- and lactate oxidase-based amperometric biosensor for lactic acid

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 79, 期 1, 页码 33-38

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/S0925-4005(01)00845-0

关键词

biosensor; Prussian Blue; lactate oxidase; lactic acid; nation; flow-injection analysis

向作者/读者索取更多资源

The properties of amperometric lactate biosensors based on Prussian Blue (PB) and lactate oxidases (LOX) of different origin (from Pediococcus sp. and Enterococcus sp. RL1) immobilised in Nafion were investigated by flow injection analysis. The operating potential was -50 mV (versus Ag/AgCl, 0.1 M KCl). A linear range in the dependence of current response upon the concentration of lactic acid was up to 0.8 mM. The detection limit of electrodes containing a high amount (6 U) of enzyme was below 1 muM. Operational stability could be maintained by exposing the electrodes to low (20 muM) concentrations of lactic acid. (C) 2001 Elsevier Science B.V. All rights reserved.