4.6 Article

Analysis of B1 field profiles and SAR values for multi-strut transverse electromagnetic RF coils in high field MRI applications

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 46, 期 10, 页码 2545-2555

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/46/10/303

关键词

-

向作者/读者索取更多资源

In this work, the B-1 field homogeneity and specific absorption rate (SAR) values were evaluated for three high-frequency (340 MHz) radio frequency coils designed for use in human magnetic resonance imaging at 8 tesla. Eight-, 16-, and 24-strut transverse electromagnetic (TEM) resonators were examined both experimentally and with the finite difference time domain numerical method. It was observed that increasing the number of TEM elements acted to lower the maximal achievable frequency of the coil and to increase the experimental complexities associated with tuning and matching, In addition, it is demonstrated from experiment and numerical analysis that the circularly polarized component of the B-1 (B-1(+)) field homogeneity in the head improved most from 8- to 16-strut coils. Numerical analysis revealed little difference in terms of SAR distribution between these coils; however, stronger tissue/coil coupling and consequently higher SAR peak values were obtained for the 8-strut case.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据