Low concentrations of phenols and amines in environmental waters and their low breakthrough volume during solid-phase extraction (SPE) hinder the detection of phenols and aromatic amines, whereas ammonia and aliphatic amines are not suitable for SPE. Pre-column derivatization to arylbenzoates and N-alkyl- or N-arylbenzamides and their GC-MS is proposed to separate and determine phenols and amines in aqueous samples in the range 0.1-100 mug l(-1) with correlation coefficients in the range 0.9910-0.9992. The limit of detection ranged from 7 to 39 ng l(-1) for most analytes (90 ng l(-1) for 2,3,6-trichlorophenol and 20 mug l(-1) for ammonia) when 80 ml of sample were preconcentrated, after derivatization, on a styrene-divinylbenzene copolymer sorbent. The developed method was applied to spiked drinking water, groundwater and river water samples, and was used to detect halo-phenols in paper mill effluents. The average recovery ranged from 96 to 110% with RSD of 4-12%. The described method is rapid and can be applied to control the water quality of environmental waters with respect to three important classes of organic pollutants and ammonia.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐