Thorium isotopic measurements on silicate rock samples with a multi-collector inductively coupled plasma mass spectrometer

期刊

ANALYST
卷 126, 期 10, 页码 1707-1710

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b101593l

关键词

-

向作者/读者索取更多资源

Chemical and mass spectrometric procedures for thorium isotopic analysis using a highly sensitive multiple-collector inductively coupled plasma mass spectrometer (MC-ICP-MS) equipped with a retardation lens were developed. They were applied to thorium isotopic measurements in volcanic rock samples. Rock samples were digested in screw-capped Savillex beakers with HF-HClO4. Thorium was then isolated from the rock matrix by a conventional anion-exchange procedure. Purified thorium was analyzed using an MC-ICP-MS (Micromass, IsoProbe) with a micro-concentric nebulizer (Cetac Aridus). The sensitivity for thorium was about 80 V ppm(-1), which enabled us to obtain a strong enough ion current to measure Th-230/Th-232 ratios with a 10 ppb solution. With these procedures, we were able to reduce the amount of thorium required for precise isotopic analysis to <10 ng. Mass-fractionation correction and Daly/Faraday multiplier gain calibration were applied using natural uranium standard solution. The procedures were applied to standard rock samples whose thorium isotopic ratios had been reported using thermal ionization mass spectrometry (TIMS). Thorium isotopic compositions measured for JB-1 and AGV-1 gave coincident values with the TIMS values. In addition, the Th-230/Th-232 radioactivity ratios of the three standard rocks including JR-2, with eruption ages of older than 350 kyr, were in equilibrium with those of U-238/Th-232, confirming the accuracy of our analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据