4.4 Article

On the effects of gating in diffusion imaging of the brain using single shot EPI

期刊

MAGNETIC RESONANCE IMAGING
卷 19, 期 8, 页码 1125-1128

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/S0730-725X(01)00415-5

关键词

diffusion tensor imaging; magnetic resonance imaging; cardiac gating; triggering; artifacts

向作者/读者索取更多资源

To reduce image artifacts in diffusion tensor imaging using single shot EPI, cardiac gating may be performed to prevent brain motion induced signal void in the DWI data. In this work the necessity of performing cardiac gating with single shot echo planar imaging has been explored using F statistics of the variance in DWI data. Peripheral measurement of the cardiac cycle has been employed because of its greater convenience, and hence use, compared to ECG gating. Four volunteers have been studied. Six different trigger delays have been evaluated in the range 18-500 ms relative the peripheral pulse wave. Brain motion was found significant in and inferior to the corpus callosum area. A trigger delay of 500 ms was found to be optimal. However, a long delay allows only for one slice per R-R interval. Therefore, a minimum trigger delay may be the best choice in terms of SNR per unit of experiment time. (C) 2001 Elsevier Science Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据