4.2 Article

Comparison of bioimpedance versus thermodilution cardiac output during cardiac surgery: Evaluation of a second-generation bioimpedance device

期刊

出版社

W B SAUNDERS CO
DOI: 10.1053/jcan.2001.26533

关键词

cardiac output; monitoring; bioimpedance; pulmonary artery catheter; cardiopulmonary bypass

向作者/读者索取更多资源

Objective: To compare a second-generation thoracic electrical bioimpedance (TEB) hemodynamic monitoring system with the clinically used pulmonary artery catheter thermodilution (TD-PAC) system. Design: Blinded, simultaneous measurements at specified key time points during surgery. Setting: University teaching hospital cardiac surgical operating rooms. Participants: Forty-seven patients undergoing primary elective coronary artery bypass surgery. Interventions: Timed cardiac output measurements by thermodilution and continuous monitoring of bioimpedance were performed. Measurements and Main Results: Cardiac index (TEB and TD-PAC) and other hemodynamic parameters were measured at 4 time points: (1) after anesthesia induction, (2) with the mediastinum open, (3) immediately after cardiopulmonary bypass, and (4) at the end of the case. Pearson's correlation and Bland-Altman analysis were carried out. Cardiac index by TEB and TD-PAC had an overall correlation of r = 0.71 (p < 0.0001). The Bland-Altman statistics showed a mean difference of -0.28 L/min/m(2) and precision of 0.67 L/min/m2. The best correlation was at time 1, and the lowest correlation was at time 4. Mediastinal opening and cardiopulmonary bypass had little or no effect on the correlation between technologies. Conclusion: TEB reporting of cardiac index during coronary artery surgery generally agreed with TD-PAC cardiac index except at the end of the case (time 4). Copyright (C) 2001 by W.B. Saunders Company.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据