期刊
JOURNAL OF ELECTROANALYTICAL CHEMISTRY
卷 512, 期 1-2, 页码 101-109出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/S0022-0728(01)00595-2
关键词
biosensors; covalent immobilisation; conducting polymers; glucose; electrochemical quartz microbalance
We have constructed and characterised a glucose sensor using glucose oxidase (GOD) covalently attached to carboxylic acid polyethyleneglycol (PEG), called (PEG-GOD). This modified enzyme was entrapped afterwards within poly(3,4-ethylenedioxythiophene) (PEDT) films electrogenerated on glassy carbon (GC) electrodes. The composite (PEG-GOD/PEDT) film is more porous than the film without enzyme (PEDT + PEG). Data from electrochemical quartz microbalance (ECQM) and pH-stat experiments indicate a good relative activity of the modified enzyme, ca. 12-15%. Amperometric measurements, using ferrocenemethanol as the redox mediator, confirms that the modified enzyme is catalytically active. The effect of film thickness was also investigated. The sensitivities were quite similar for modified-GOD electrodes (ca. 3 mA cm (-2) M (-1)) and unmodified-GOD electrodes (ca. 2.7 mA cm -(2) M (-1)) but a better stability was obtained with modified PEG-GOD electrodes. (C) 2001 Elsevier Science B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据