4.8 Article

Electrochemical enzyme immunoassays on microchip platforms

期刊

ANALYTICAL CHEMISTRY
卷 73, 期 21, 页码 5323-5327

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac010808h

关键词

-

资金

  1. PHS HHS [R01 14549-02] Funding Source: Medline

向作者/读者索取更多资源

A microfluidic device for conducting electrochemical enzyme immunoassays is described. The new lab-on-a-chip protocol integrates precolumn reactions of alkaline phosphatase-labeled antibody (anti-mouse IgG) with the antigen (mouse IgG), followed by electrophoretic separation of the free antibody and antibody-antigen complex. The separation is followed by a postcolumn reaction of the enzyme tracer with the 4-aminophenyl phosphate substrate and a downstream amperometric detection of the liberated 4-aminophenol product. Factors influencing the reaction, separation, and detection processes were optimized, and the analytical performance was characterized. An applied field strength of 256 V/cm results in free antibody and antibody-antigen complex migration times of 125 and 340 s, respectively. A remarkably low detection limit of 2.5 x 10(-16) g/mL (1.7 x 10(-18) M) is obtained for the mouse IgG model analyte. Such combination of a complete integrated immunoassay, an attractive analytical performance, and the distinct miniaturization/portability advantages of electrochemical microsystems offers considerable promise for designing self-contained and disposable chips for decentralized clinical diagnostics or onsite environmental testing.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据