In this article, we describe the use of pH-responsive hydrogels as matrices for the immobilization of two enzymes, glucose oxidase (GOx) and glutamate oxidase (GlutOx). Spherical hydrogel beads were prepared by inverse suspension polymerization and the enzymes were immobilized by either physical entrapment or covalent immobilization within or on the hydrogel surface. Packed-bed bioreactors were prepared containing the bioactive hydrogels and these incorporated into flow injection (Fl) systems for the quantitation of glucose and monosodium glutamate (MSG) respectively. The Fl amperometric detector comprised a microfabricated Interdigitated array within a thin-layer flow cell. For the Fl manifold incorporating immobilized GOx, glucose response curves were found to be linear over the concentration range 1.8-280 my dL(-1) (0.1-15.5 mM) with a detection limit of 1.4 mg dL(-1) (0.08 mM). Up to 20 samples can be manually analyzed per hour, with the hydrogel-GOx bioreactor exhibiting good within-day (0-19%) precision. The optimized Fl manifold for MSG quantitation yielded a linear response range of up to 135 mg dL(-1) (8 mM) with a detection limit of 3.38 mg dL(-1) (0.2 mM) and a throughput of 30 samples h(-1). Analysis of commercially produced soup samples gave a within-day precision of 3.6%. Bioreactors containing these two physically entrapped enzymes retained > 60% of their initial activities after a storage period of up to 1 year. (C) 2001 John Wiley & Sons, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据