4.7 Article

Lack of mutagenic and co-mutagenic effects of magnetic fields during magnetic resonance Imaging

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 14, 期 6, 页码 779-788

出版社

WILEY-BLACKWELL
DOI: 10.1002/jmri.10010

关键词

mutagenicity; co-mutagenicity; magnetic resonance; biological effects; magnetic fields; Salmonella typhimurium

向作者/读者索取更多资源

Mutagenic and co-mutagenic effects of static, pulsed bipolar gradient, and high-frequency magnetic fields, as well as combinations of them, were examined using the Ames test. The Ames test using Salmonella typhimurium bacteria, wild-type strain RTA, preincubation assay, without metabolic activation, was performed. All combinations of magnetic fields were tested with and without co-exposure to N-methyl-N'-nitro-N-nitrosoguanidine and benzo[a]pyrene-4,5-oxide, ethylene oxide, carboplatin, or cisplatin. As expected, chemical mutagens caused a clear-cut increase of the revertants in the Ames test. However, neither the static fields nor a combination of a static magnetic field with the time-varying bipolar gradient field or a pulsed high-frequency magnetic field caused an alteration in the number of revertants in the Ames test. No co-mutagenic effect of any magnetic field combination was observed. In conclusion, magnetic fields used during clinical magnetic resonance imaging (MRI) were neither mutagenic nor co-mutagenic. (C) 2001 Wiley-Liss, Inc.