Objective data alignment and chemometric analysis of comprehensive two-dimensional separations with run-to-run peak shifting on both dimensions

期刊

ANALYTICAL CHEMISTRY
卷 73, 期 24, 页码 5833-5840

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac010656q

关键词

-

向作者/读者索取更多资源

Data from comprehensive two-dimensional (2-D) separation techniques, such as comprehensive 2-D gas chromatography (GC x GC), liquid chromatography/liquid chromatography (LC x LQ and liquid chromatography/capillary electrophoresis (LC x CE) can be readily analyzed by various Chemometric methods to increase chemical analysis capabilities. A retention time alignment, preprocessing method is presented that objectively corrects for run-to-run retention time variations on both separation dimensions of comprehensive 2-D separations prior to application of chemometric data analysis algorithms. The 2-D alignment method corrects for run-to-run shifting of a sample data matrix relative to a standard data matrix on both separation time axes in an independent, stepwise fashion. After 2-D alignment, the generalized rank annihilation method (GRAM) is successfully applied, substantiating the performance of the alignment method. The alignment method should have important implications, because most 2-D separation techniques exhibit, in the context of chemometric data analysis, considerable run-to-run retention time shifting on both dimensions. Even when there are only three to four points/peak, that is, with three to four separations on the second dimension (column 2) per peak width from the first dimension (column 1), the 2-D alignment coupled with GRAM provides dependable analyte peak identification capabilities and adequate quantitative precision for unresolved analyte peaks. Thus, the 2-D alignment algorithm is applicable to lower data density conditions, which broadens the scope of chemometric analysis to high-speed 2-D separations.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据