4.6 Article

Gross Trunnion Failure After Primary Total Hip Arthroplasty

期刊

JOURNAL OF ARTHROPLASTY
卷 30, 期 4, 页码 641-648

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2014.11.023

关键词

trunnion; total hip arthroplasty; fracture; trunnionosis; wear; failure

向作者/读者索取更多资源

Unfavorable outcomes from trunnion fretting and corrosion damage have been reported in the literature, gross failures of tapers in primary total hip arthroplasties have been less frequently reported. We report on 5 patients, who presented with gross trunnion failures of modular metal-on-polyethylene or ceramic-on-polyethylene bearings from 5 implant manufacturers, all necessitating revision surgery. None of these patients had an antecedent history of trauma, and the majority presented with pain or instability. No common factor was identified that may be predictive of these type of failures. Since there were 5 different stem designs, this suggests that it may be a rare generic phenomenon occurring withmultiple designs. Currently, further investigations are necessary, including retrieval analysis, to identify risk factors that may predispose to such failures. (C) 2014 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据