4.7 Article

Water glass bonding for micro-total analysis system

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 81, 期 2-3, 页码 187-195

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/S0925-4005(01)00951-0

关键词

water glass bonding; micro-chip; micro-chemical analysis

向作者/读者索取更多资源

We used water glass solution for bonding two glass wafers on which electrodes and a micro-channel were fabricated. This bonding technique makes possible to bond two glass wafers without concerning the problem of surface roughness. We evaluated many properties of the water glass bonding for using micro-chemical analysis, such as, effect of bond temperature on bond strength, thickness of bonding layer dependence on applied pressure during bonding, and so on. Because of high content rate of sodium in water glass, we evaluated cleaning effect to decrease sodium dissolution from bonding layer. Additionally, we studied the effect of the pH of solutions on bond strength. By using this technology, we made a micro-electrochemical chip as one of applications and identified an oxygen evolution current without break down of the chip. We show this bonding technique is very useful to make a micro-chip. (C) 2002 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据