4.6 Article

Serum Metal Ions with a Titanium Modular Neck Total Hip Replacement System

期刊

JOURNAL OF ARTHROPLASTY
卷 30, 期 10, 页码 1781-1786

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2015.04.040

关键词

arthroplasty; hip; modular; titanium; ions

资金

  1. Wright Medical Technologies (Memphis, Tennessee)

向作者/读者索取更多资源

The goal of this study is to evaluate serum levels of chromium(Cr), cobalt (Co), and titanium(Ti) within the first two years following total hip arthroplasty using a Ti modular neck system. Twenty-five patients were randomized to a metal-on-metal (MoM) bearing with an all CoCr shell, and the remaining 25 received a metal-on-polyethylene (MoP) with a Ti shell. Serum levels demonstrated increases for Cr, Co, and Ti at 1 year (P <.001). MoM had similar Ti levels to MoP hips at 1 year (P = 0.11) but lower at 2 years (P = 0.03). Results suggest that the passive corrosion (i.e., chemical, pitting, and crevice corrosion) of exposed non-articular metal surfaces may be a greater source of ions than the neck-stem or head-neck interfaces. (c) 2015 Elsevier Inc. All rights reserved.