期刊
JOURNAL OF CHROMATOGRAPHY A
卷 947, 期 2, 页码 155-165出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/S0021-9673(01)01609-0
关键词
solid-phase microextraction; water analysis; environmental analysis; wood; chlorophenols; pentachlorophenol
A solid-phase microextraction method has been developed for the determination of 19 chlorophenols (CPs) in environmental samples. The analytical procedure involves direct sampling of CPs from water using solid-phase microextraction (SPME) and determination by liquid chromatography with electrochemical detection (LC-ED). Three kinds of fibre 150 mum carbowax-templated resin (CW-TPR), 60 mum polydimethylsiloxane-divinylbenzene (PDMS-DVB) and 85 mum polyacrylate (PA)] were evaluated for the analysis of CPs. Of these fibres, CW-TPR is the most suitable for the determination of CPs in water. Optimal conditions for both desorption and absorption SPME processes, such as composition of the desorption solvent (water-acetonitrile-methanol, 20:30.50) and desorption time (5 min), extraction time (50 min) and temperature (40 degreesC) as well as pH (3.5) and ionic strength (6 g NaCl) were established. The precision of the SPME-LC-ED method gave relative standard deviations (RSDs) of between 4 and 11%. The method was linear over three to four orders of magnitude and the detection limits, from 3 to 8 ng l(-1), were lower than the European Community legislation limits for drinking water. The method was applied to the analysis of CPs in drinking water and wood samples. (C) 2002 Elsevier Science B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐