4.0 Article

Comparison of cyclic fatigue resistance of novel nickel-titanium rotary instruments

期刊

AUSTRALIAN ENDODONTIC JOURNAL
卷 41, 期 1, 页码 24-28

出版社

WILEY-BLACKWELL
DOI: 10.1111/aej.12067

关键词

cyclic fatigue; heat treatment; nickel-titanium files; ProTaper Next; rotary

向作者/读者索取更多资源

New files (ProTaper Next/HyFlex/OneShape) are made from novel nickel-titanium (NiTi) alloys/treatments. The purpose of this study was to compare the cyclic fatigue resistance of these new instruments with that of Revo-S instruments. Four groups of 20 NiTi endodontic instruments were tested in steel canals with a 3mm radius and a 60 degrees angle of curvature. The cyclic fatigue of the following NiTi instruments with a tip size 25 and 0.06 taper that were manufactured with different alloys was tested: ProTaper Next X2 (M-Wire), OneShape (conventional NiTi), Revo-S Shaping Universal (conventional NiTi) and HyFlex 25/0.6 (controlled memory NiTi wire). A one-way anova and post-hoc Tukey's test (=0.05) revealed that the HyFlex files had the highest fatigue resistance and the Revo-S had the least fatigue resistance among the groups (P<0.001).