4.3 Article

Comparison of the axial resolution of practical Nipkow-disk confocal fluorescence microscopy with that of multifocal multiphoton microscopy: theory and experiment

期刊

JOURNAL OF MICROSCOPY
卷 206, 期 -, 页码 24-32

出版社

WILEY-BLACKWELL
DOI: 10.1046/j.1365-2818.2002.01001.x

关键词

axial resolution; confocal microscopy; multifocal multiphoton microscopy

向作者/读者索取更多资源

We compare the axial sectioning capability of multifocal confocal and multifocal multiphoton microscopy in theory and in experiment, with particular emphasis on the background arising from the cross-talk between adjacent imaging channels. We demonstrate that a time-multiplexed non-linear excitation microscope exhibits significantly less background and therefore a superior axial resolution as compared to a multifocal single-photon confocal. system. The background becomes irrelevant for thin (< 15 mu m) and sparse fluorescent samples, in which case the confocal parallelized system exhibits similar or slightly better sectioning behaviour due to its shorter excitation wavelength. Theoretical and experimental axial responses of practically implemented microscopes are given.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据