4.6 Article

A new paradigm for the closed-loop intraoperative administration of analgesics in humans

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 49, 期 4, 页码 289-299

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/10.991156

关键词

alfentanil; analgesia; artifact tolerant control; closed-loop control; model-predictive control (MPC)

向作者/读者索取更多资源

We present a new paradigm for the closed-loop administration of analgesics during general anesthesia. The manipulated variable in the control system is the infusion rate of the opiate alfentanil, administered intravenously through a computer-controlled infusion pump (CCIP). The outputs to be controlled are the patient's mean arterial pressure (MAP) and the drug concentration in the plasma. Maintaining MAP within appropriate ranges provides optimal treatment of the patient's reactions to surgical stimuli. Maintaining plasma drug concentrations close to a reference value specified by the anesthesiologist allows to titrate analgesic administration to qualitative clinical end-points of insufficient analgesia. MAP is acquired invasively through a catheter cannula. Since plasma drug concentrations cannot be measured on-line, they are estimated via a pharmacokinetic model. We describe an explicit model-predictive controller which achieves the above-mentioned objectives. An upper constraint on drug concentrations is maintained to avoid overdosing. Constraints on the MAP are introduced to trigger a prompt controller reaction during hypertensive and hypotensive periods. Measurement artifacts in the MAP signal are rejected to prevent harmful misbehavior of the controller. We discuss the results of the clinical validation of the controller on humans.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据