4.5 Article

Origin and minimization of residual motion-related artifacts in navigator-corrected segmented diffusion-weighted EPI of the human brain

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 47, 期 4, 页码 818-822

出版社

WILEY
DOI: 10.1002/mrm.10102

关键词

diffusion-weighted imaging; navigator correction; brain motion; cardiac pulsation; artifacts

资金

  1. NCRR NIH HHS [P41 RR15241] Funding Source: Medline
  2. NIBIB NIH HHS [P41 EB015909] Funding Source: Medline
  3. NICHD NIH HHS [R01 HD37931] Funding Source: Medline

向作者/读者索取更多资源

Motion sensitivity in diffusion-weighted imaging (DWI) can be effectively suppressed using single-shot echo-planar imaging (EPI). However, segmented (multishot) EPI is often used to increase resolution and reduce spatial distortions, which in turn increases susceptibility to brain motion. The sources of these residual motion artifacts in navigator-echo-corrected segmented EPI images of the brain were investigated. The results indicate that the dominant source of these artifacts is cardiac pulsation with occasional involuntary movement of the subject. The relationship between the cardiac cycle and motion artifacts shows that optimum timing for the data acquisition is possible. In addition it is shown that the effects of involuntary motion can be removed by swapping k-space data between redundant datasets. (C) 2002 Wiley-Liss, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据