期刊
SENSORS AND ACTUATORS A-PHYSICAL
卷 97-8, 期 -, 页码 280-288出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/S0924-4247(01)00847-0
关键词
RF microcoil; high quality factor; detection methods; nuclear magnetic resonance
We present the design, fabrication and test of high-Q factor radiofrequency planar microcoils for nuclear magnetic resonance (NMR) spectroscopy in small volume samples. The coils are fabricated on glass wafers using high-aspect ratio SU-8 photoepoxy and copper electroplating. On-wafer electrical characterization shows quality factors up to 40 at 800 MHz. A 500 mum diameter microcoil with a measured quality factor of 24 at 300 MHz is mounted on a printed circuit board and electrically contacted using aluminum wire bonding. This probe is inserted in a 7 T-superconducting magnet, and a H-1-NMR spectrum of 160 nl ethylbenzene contained in a capillary placed over the microcoil is acquired in a single scan. This work is an important step towards the integration of NMR detection into micro-total analysis s' stems (muTAS). (C) 2002 Elsevier Science B.V. All rights reserved.