4.4 Article

A versatile drop foot stimulator for research applications

期刊

MEDICAL ENGINEERING & PHYSICS
卷 24, 期 3, 页码 237-242

出版社

ELSEVIER SCI LTD
DOI: 10.1016/S1350-4533(02)00011-5

关键词

stimulator; tibialis anterior; drop foot; common peroneal; gait correction

向作者/读者索取更多资源

Drop Foot Stimulators are used to correct hemiplegic drop foot by synchronising the application of Functional Electrical Stimulation (FES) of the Common Peroneal Nerve (CPN) to the swing phase of the gait cycle. A research Drop Foot Stimulator (DFS) has been developed with a very flexible architecture to enable the investigation of a variety of gait-correction strategies. The portable unit has been carefully designed to optimise functionality while keeping its size and power consumption to a minimum. The device has two channels of stimulation, with all parameters of stimulation for each channel independently programmable. Four analogue and four digital sensor input channels are provided with a wide variety of sensor types possible. A microcontroller core is utilised to enable the implementation of different control algorithms. A PC-based user interface enables easy programming of the system configuration. (C) 2002 IPEM. Published by Elsevier Science Ltd. All rights reserved.