An experimental comparison of detector performance for computed radiography systems

期刊

MEDICAL PHYSICS
卷 29, 期 4, 页码 447-459

出版社

WILEY
DOI: 10.1118/1.1449873

关键词

digital radiography; computed radiography (CR); image quality; modulation transfer function (MTF); detective quantum efficiency (DQE)

向作者/读者索取更多资源

The intrinsic resolution, noise, and signal-to-noise transfer characteristics of five commercial digital computed radiography (CR) systems were compared using identical experimental methods. The reader/screen combinations evaluated were Agfa ADC-Compact/MD-10, Agfa ADC-Compact/MD-30, Agfa ADC-Solo/MD-10, Agfa ADC-Solo/MD-30, Lumisys CR-2000/MD-10, Fuji FCR9501(HQ)/ST-Va, Kodak CR-400/GP-25, and Kodak CR-400/HR. Measurements were made at 70 and 115 kVp with 19 mm added aluminum filtration. The presampled modulation transfer functions (MTFs) of the systems were measured using an edge method. The noise power spectra (NPS) were determined by 2D Fourier analysis of uniformly exposed radiographs. The frequency-dependent detective quantum efficiencies (DQEs) were computed from the MTF, NPS, exposure measurements, and computational estimates of the ideal signal-to-noise ratios. Using 70 kVp and 0.1-0.12 mm pixel sizes, spatial frequencies of 2.1, 2.0, 2.2, 1.9, 2.0, 2.0, 23, 2.3, and 3.5 cycles/mm were measured at 0.2 MTF for the eight reader/screen combinations, respectively. Using 70 kVp, 7.74 X 10(-8) C/kg (0.3 mR), and 0.1-0.12 mm pixel sizes, DQE(0.15) values of 20.3%, 22.9%, 24.6%, 28.6%, 22.2%, 30.0%, 29.5%, and 17.3% were obtained for the eight combinations, respectively. The corresponding values at 115 kVp were 15.9%, 18.5%, 21.5%, 21.8%, 15.3%, 23.1%, 22.3%, and 13.8%, respectively. The findings of the study demonstrate the pixel size, orientation. beam quality, screen, and reader dependencies of image quality in CR systems. The physical performance of the systems having standard-resolution screens demonstrated similar resolution performance but more notable variations in DQE. The one high-resolution screen tested had reduced DQE and increased MTF at high frequencies. (C) 2002 American Association of physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据