4.6 Article

Cardiac output measurement by pulse dye densitometry: a comparison with the Fick's principle and thermodilution method

期刊

INTENSIVE CARE MEDICINE
卷 28, 期 4, 页码 399-405

出版社

SPRINGER-VERLAG
DOI: 10.1007/s00134-002-1252-3

关键词

cardiac output; thermodilution technique; Fick technique; dye dilution technique; postoperative; cardiopulmonary bypass

向作者/读者索取更多资源

Objective: To evaluate the agreement between cardiac output (CO) measurements obtained by a new dye dilution technique using pulse dye densitometry (PDD) and thermodilution (TD) and the direct Fick method (F). Design and setting: Prospective clinical study in a university hospital, cardiac surgery intensive care unit. Patients: Fifty-eight cardiac surgery patients after admission to the intensive care unit (six were excluded due to a low pulse signal quality using the PDD method). Measurements and results: Mean CO was 5.3+/-1.81/min for PDD, 5.7+/-1.68 l/min for TD, and 6.16+/-1.66 l/min for F. There was a good correlation between PDD and TD (r(2)=0.93) and between PDD and F (r(2)=0.77). Bias and precision between PDD and TD were -0.39+/-0.5 l/min and -0.69+/-0.85 l/min between PDD and F. In general, PDD determined lower CO values than TD and F. Especially in patients with CO below 5 l/min PDD underestimated CO in comparison to TD and F (bias and precision: -0.51+/-0.40 l/min and -0.83+/-1.0 l/min). Conclusion: Comparison between PDD and TD showed good agreement for the normal to high CO range. However, agreement was poor in patients with low CO. In the latter patient group PDD showed relevant underestimation of CO compared to TD and F. Due to these Limitations PDD cannot entirely replace the pulmonary artery catheter for CO determination.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据