4.3 Article

Patellar instability following total knee replacement - the dangers of constant design evolution

期刊

KNEE
卷 9, 期 2, 页码 151-153

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0968-0160(01)00141-7

关键词

total knee replacement; patellar instability; prosthetic design

向作者/读者索取更多资源

Patellar instability following total knee replacement although infrequent can cause disabling symptoms. We report a case of symptomatic patellar instability caused by insertion of left femoral component into a right knee. We highlight that all members of the surgical and theatre team need to be familiar both with the implant equipment and of any recent design modifications, particularly if they have been used to using one system for a long time, in order to prevent surgical errors. In addition this possibility should be born in mind by surgeons looking for causes of post-operative patellar instability. (C) 2002 Elsevier Science B.V. All rights reserved.