The design and characterization of instrumentation for application in evaluating the neural signal recording properties of probe-type microelectrodes, micromachined from silicon, are reported. Key aspects include the close matching of gain and frequency response between channels (better than 1%), flexibility in signal conditioning options, the ability to operate with a wide range of (microelectrode) recording site dimensions (4 mum x 4 mum to 50 mum x 50 mum), and hence impedances, and the facility to monitor and store instrumentation settings on computer along with the recorded signals. Noise levels ranged from 3.7 muV rms for a 50 mum site, to 11.7 muV rms for a 4 mum site, measured in saline. Close matching between channels was required to enable comparisons between different sites and different probes to be made with confidence; however, the instrumentation could be readily applied to less demanding applications.