期刊
PHYSIOLOGICAL MEASUREMENT
卷 23, 期 2, 页码 437-448出版社
IOP PUBLISHING LTD
DOI: 10.1088/0967-3334/23/2/317
关键词
neurophysiology instrumentation; microelectrode; neurotechnology
The design and characterization of instrumentation for application in evaluating the neural signal recording properties of probe-type microelectrodes, micromachined from silicon, are reported. Key aspects include the close matching of gain and frequency response between channels (better than 1%), flexibility in signal conditioning options, the ability to operate with a wide range of (microelectrode) recording site dimensions (4 mum x 4 mum to 50 mum x 50 mum), and hence impedances, and the facility to monitor and store instrumentation settings on computer along with the recorded signals. Noise levels ranged from 3.7 muV rms for a 50 mum site, to 11.7 muV rms for a 4 mum site, measured in saline. Close matching between channels was required to enable comparisons between different sites and different probes to be made with confidence; however, the instrumentation could be readily applied to less demanding applications.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据