IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 49, 期 5, 页码 472-483出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/10.995686
关键词
calibration; four-terminal; heart; in-vivo; leakage current; myocardium; pig; plunge probe; probe constant; resistivity; stray capacitance
资金
- NHLBI NIH HHS [HL56143] Funding Source: Medline
We used a four-terminal plunge probe to measure myocardial resistivity in two directions at three sites from the epicardial surface of eight open-chest pigs in-vivo at eight frequencies ranging from I Hz to I MHz. We calibrated the plunge probe to minimize the error due to stray capacitance between the measured subject and ground. We calibrated the probe in saline solutions contained in a metal cup situated near the heart that had an electrical connection to the pig's heart. The mean of the measured myocardial resistivity was 319 Omega . cm at I Hz down to 166 Omega . cm at I MHz. Statistical analysis showed the measured myocardial resistivity of two out of eight pigs was significantly different from that of other pigs. The myocardial resistivity measured with the resistivity probe oriented along and across the epicardial fiber direction was significantly different at only one out of the eight frequencies. There was no significant difference in the myocardial resistivity measured at different sites.