Preconcentration of Cd(II), Cu(II), Mn(II), Ni(II), Pb(II) and Zn(II) in saline matrices on Amberlite XAD-4 resins coated with ammonium pyrrolidine dithiocarbamate (APDC) and piperidine dithiocarbamate (pipDTC) and subsequent determination by inductively coupled plasma atomic emission spectrometry were studied. Parameters such as effect of pH, effect of HNO3 concentration on elution of metals from resin were studied. The results show that Amberlite XAD-4 coated with APDC was more efficient in the recovery of metal ions compared with Amberlite XAD-4 coated with pipDTC, in the concentration range of 0.1-200 mug l(-1), for 1 g of Amberlite XAD-4 coated resin. The detection limits for Cd(II), Cu(II), Mn(II), Ni(II), Pb(II), Zn(II) are 0.1, 0.4, 0.3, 0.4, 0.6, 0.5 mug l(-1), respectively, for resin coated with APDC and 0.7, 1.0, 0.8, 0.9, 1.7 and 1.2 mug l(-1) for resin coated with pipDTC. The effect of diverse ions on the determination of aforesaid metals was studied. The method was applied for the determination of trace metal ions in artificial sea water and natural water samples. The results were compared with extraction AAS method. (C) 2002 Elsevier Science B.V. All rights reserved.