Realization of matching conditions for high-resolution spectrometers

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0168-9002(01)01970-2

关键词

magnetic spectrometer; high resolution; large dispersion; kinematic correction; focus matching; lateral dispersion matching; angular dispersion matching

向作者/读者索取更多资源

For precise measurements of nuclear-reaction spectra using a beam from an accelerator, a high-resolution magnetic spectrometer is a powerful tool. The full capability of a magnetic spectrometer, however, can be achieved only if the characteristics of the beam coming from the accelerator are matched to those required by the spectrometer by properly, adjusting the beam line conditions. The matching methods, lateral dispersion matching, focus matching and also the kinematic correction compensate the spectrum line-broadening effects caused by the beam momentum spread and reaction kinematics. In addition, angular dispersion matching should be performed if good resolution of the scattering angle is important. Diagnostic methods developed to realize these matching conditions for the spectrometers K600 at IUCF and Grand Raiden at RCNP are presented. (C) 2001 Elsevier Science B.V. All rights reserved.