期刊
JOURNAL OF ARTHROPLASTY
卷 17, 期 4, 页码 167-170出版社
CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1054/arth.2002.32465
关键词
cement; acetabular; liner; cementless; shell
类别
A patient presenting with a Secure cementless acetabular component and with femoral head penetration through the polyethylene liner is a common clinical problem. Cementing a new liner into the fixed shell is one option. We evaluated this option in a clinical series of 17 cases and with a preliminary mechanical study. In the I Clinical failure (5.9%), the failure occurred at the cement-liner interface, The most important variable in optimizing the mechanical strength of the cemented liner construct was adequate preparation of the cement-liner interface. This approach to treating the patient with a fixed cementless shell and a worn polyethylene liner can provide a durable construct with minimal morbidity.