4.4 Article

Effectiveness and relevance of MR acceptance testing: results of an 8 year audit

期刊

BRITISH JOURNAL OF RADIOLOGY
卷 75, 期 894, 页码 523-531

出版社

BRITISH INST RADIOLOGY
DOI: 10.1259/bjr.75.894.750523

关键词

-

向作者/读者索取更多资源

The effectiveness and relevance of independent acceptance testing was assessed by means of an audit of acceptance procedures for 17 MRI systems, with field strengths in the range 0.5-1.5 T, acquired over 8 years. Signal-to-noise ratio and geometric linearity were found to be the image quality parameters most likely to fall below acceptable or expected standards. These received confirmed successful corrective action in 69% of instances. Non-uniformity, ghosting and poor fat suppression were the next most common non-compliant parameters, but yielded less satisfactory outcomes. Spatial resolution was not found to be a sensitive parameter in determining acceptability. 49% of all non-compliant parameters received verifiable corrective attention. A schedule of actual acceptance criteria is presented and shown to be reasonable. Parameter failure rates were shown not to have improved with time. A safety audit of 11 of the installations revealed the most common failings to be inadequate suite layout and poor use of signs. The mean number of safety issues per installation identified as requiring attention was 5, from a questionnaire of 100 points. A number of anecdotal errors and omissions are reported. The data support the importance of an appropriate acceptance procedure for new clinical MRI equipment and for the involvement of a suitably qualified safety adviser on the project team from the outset.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据