4.6 Article

Determination of arsenic by hydride generation coupled to time-of-flight mass spectrometry with a gas sampling glow discharge

期刊

SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
卷 57, 期 6, 页码 1009-1023

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0584-8547(02)00026-5

关键词

hydride generation; arsenic determination; time-of-flight mass spectrometry; gas sampling glow discharge

向作者/读者索取更多资源

Hydride generation has been used with a gas-sampling glow discharge (GSGD) and time-of-flight mass spectrometry (TOFMS) for the determination of arsenic in solution. Helium, neon, hydrogen, and argon glow discharges have been successfully generated and characterized. Current-pressure-voltage curves were generated for each discharge in the presence and absence of hydride generation. The arsenic detection limit for each of the discharges was found to be 0.60 (HeGSGD), 3.8 (NeGSGD) and 6.4 ppb (H(2)GSGD). The HeGSGD was found to be the most attractive source for arsenic determination due to the lower detection limit, higher sensitivity and greater stability. The figures of merit for these discharges were also compared to those obtained with hydride generation-inductively coupled plasma TOFMS. Noise power spectra obtained for the neon GSGD indicated that no discernible discrete-frequency (whistle-noise) components were present in the analyte signal. (C) 2002 Elsevier Science B.V. All rights reserved.