4.3 Article Proceedings Paper

The Esophageal Detector Bulb in the aeromedical setting

期刊

JOURNAL OF EMERGENCY MEDICINE
卷 23, 期 1, 页码 51-55

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/S0736-4679(02)00462-6

关键词

intubation; emergency medical services; air ambulances; esophageal detector device; aeromedicine

向作者/读者索取更多资源

To determine the accuracy of the Esophageal Detector Bulb (EDB) in the aeromedical setting, we conducted a prospective, observational study of all intubated patients transported by an aeromedical program over two years. Flight personnel recorded the results of the EDB, clinical examination, pulse oximetry, and capnography (ETCO2). Endotracheal tube position was confirmed by prehospital ETCO2 or by the receiving emergency physician. There were 104 EDB assessments in 53 patients. The EDB correctly identified four of five esophageal intubations and 96 of 99 tracheal intubations. The sensitivity and specificity of the EDB in the detection of an esophageal intubation were 80% (95% CI, 38-96%) and 97% (95% CI, 92-99%), respectively, and the overall accuracy was 96% (95% CI, 90-98%). The EDB augments the ability, of an aeromedical crew to determine endotracheal tube position, but its results must be carefully interpreted in the context of other available means of confirmation of endotracheal tube position. (C) 2002 Elsevier Science Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据