4.6 Article

Demonstration of a high-performance pulsed optically pumped Rb clock based on a compact magnetron-type microwave cavity

期刊

JOURNAL OF APPLIED PHYSICS
卷 117, 期 10, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.4914493

关键词

-

资金

  1. Swiss National Science Foundation (SNSF) [140712]
  2. European Metrology Research Programme (EMRP) [IND55-Mclocks]
  3. EMRP within EURAMET
  4. European Union

向作者/读者索取更多资源

We demonstrate a high-performance pulsed optically pumped (POP) Rb vapor-cell clock based on a magnetron-type microwave cavity of only 44 cm(3) external volume. Using optical detection, an unprecedented 35% contrast of the Ramsey signal has been obtained. Both the signal-to-noise ratio (of 30 000) and the estimated shot-noise limit of 1.7 x 10(-14) tau(-1/2) are at the same level as those found with a bigger cylindrical TE011 cavity (100 cm 3 inner volume) and are sufficient for achieving excellent clock stability. Rabi oscillations are measured and indicate a sufficiently uniform microwave magnetic field distribution inside the cavity. The instability sources for the POP clock's performance are analyzed. A short-term stability of 2.1 x 10(-13) tau(-1/2) is demonstrated which is consistent with the noise budget. (C) 2015 AIP Publishing LLC.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据