4.7 Article

Glucose biosensors prepared by electropolymerization of p-chlorophenylamine with and without Nafion

期刊

ANALYTICA CHIMICA ACTA
卷 463, 期 2, 页码 239-247

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0003-2670(02)00430-0

关键词

glucose oxidase; biosensor; nafion; poly(p-chlorophenylamide)

向作者/读者索取更多资源

Two different glucose biosensors for the amperometric determination of glucose, based on poly (p-chlorophenylamide) (PCPA) and bilayer film of PCPA and Nafion (PCPA/Nafion), are successfully developed. These two biosensors show linear amperometric responses to glucose ranging from 2.0 x 10(-4) to 3.5 x 10(-2) mol(-1) and 5.0 x 10(-4) to 7.5 x 10(-2) mol l(-1), respectively, with the same correlation coefficient of 0.9988, Effects of polymerization potential and polymerization time on the performance of enzyme sensors are studied, It is found that PCPA, as a non-conducting polymer, can largely reduce the influence of electroactive interferents. Introduction of inner Nafion membrane not only further eliminates the influence of ascorbic acid on the sensor response but also increases electrode stability. (C) 2002 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。