4.6 Article

Method based on solid-phase microextraction-high-performance liquid chromatography with UV and electrochemical detection to determine estrogenic compounds in water samples

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 964, 期 1-2, 页码 153-160

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0021-9673(02)00694-5

关键词

water analysis; environmental analysis; endocrine disruptors; estrogenic compounds

向作者/读者索取更多资源

We determined a group of estrogenic compounds by solid-phase microextraction (SPME) coupled to high-performance liquid chromatography (HPLC) with both ultraviolet (UV) and electrochemical detection (ED). A modified liquid chromatograph was used. Polyacrylate fibers (85 km) were used to extract the analytes from the aqueous samples. Dynamic and static modes of desorption were compared and the variables affecting both absorption and desorption processes in SPME-HPLC were optimized. Static desorption gave the best recoveries and peak shapes. The performance of the SPME-HPLC-UV-ED method was checked with river water and wastewater. The method enabled estrogenic compounds to be determined at low-mug l(-1) levels in real water samples. Limits of detection were between 0.3 and 1.1 mug l(-1) using UV detection and between 0.06 and 0.08 mug l(-1) using ED. P-Estradiol was found in samples from a wastewater treatment plant at concentrations between 1.9 and 2.2 mug l(-1). (C) 2002 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据