期刊
NEUROIMAGE
卷 16, 期 4, 页码 1151-1155出版社
ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1006/nimg.2002.1119
关键词
-
A new approach to reduce gradient acoustic noise levels in EPI experiments is presented. Using multichannel RF receive coils, combined. With SENSE data Acquisition and reconstruction, gradient slew-rates in single-shot EPI were reduced fourfold for rate-2 and ninefold for rate-3 SENSE. Multislice EPI experiments Were performed on three different scanner platforms. With 3.4 min in-plane resolution, measuring 6 slices per second (1 slices with 2000 ms TR), this resulted in average sound pressure level reductions of 11.3 dB(A) and 16.5 dB(A) for rate-2 and rate-3 SENSE, respectively. BOLD fMRI experiments, using visually paced finger-tapping paradigms, showed no detrimental effect of the acoustic noise reduction strategy con temporal noise levels and t scores.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐