4.4 Article

Effectiveness of capillary electrophoresis using fluorescent-labeled primers in detecting T-cell receptor γ gene rearrangements

期刊

JOURNAL OF MOLECULAR DIAGNOSTICS
卷 4, 期 3, 页码 137-143

出版社

AMER SOC INVESTIGATIVE PATHOLOGY, INC
DOI: 10.1016/S1525-1578(10)60694-0

关键词

-

向作者/读者索取更多资源

We describe the use of fluorescent-labeled primers to analyze T-cell receptor gamma gene rearrangements (TCRgammaGR) using capillary electrophoresis in the ABI Prism 310 Genetic Analyzer. We also compare die performance with denaturing gradient gel electrophoresis (DGGE). in a single multiplex polymerase chain reaction (PCR) we amplified TCRgammaGR with primers for all known groups of variable region genes, and joining region genes described in lymphoid neoplasms. Ten reactive samples, followed by five cell lines and 25 tumor samples with 41 individual TCRgammaGR (due to many biallelic rearrangements) previously identified by DGGE, were analyzed to validate the technique. The capillary electrophoresis. protocol has 92% concordance for both TCR clonal status (23 of 25) and 95% concordance in the number of individual TCRgammaGR (38 of 41) identified by DGGE. The reproducible sensitivity for detecting TCRgammaGR diluted in reactive lymphoid DNA is 2% in clinical applications. Discrimination of predominant rearrangements requires a minimum ratio of two times the height of the normal distribution of polyclonal peaks. Capillary electrophoresis can provide results within 60 minutes for each specimen after PCR is complete. Capillary electrophoresis provides a faster result than sequence-based separation methods and gives an archival electronic record. Fluorescent labeling allows the identification of both the variable and joining gene segments used in a TCRgammaGR. The effectiveness of capillary electrophoresis is similar to DGGE.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据