4.6 Article

Measurement of stand-sit and sit-stand transitions using a miniature gyroscope and its application in fall risk evaluation in the elderly

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 49, 期 8, 页码 843-851

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2002.800763

关键词

ambulatory system; fall risk assessment in elderly people; gyroscope; postural transition and successive transition; wavelet transform

向作者/读者索取更多资源

A new method of evaluating the characteristics of postural transition (PT) and their correlation with falling risk in elderly people is described. The time of sit-to-stand and stand-to-sit transitions and their duration were measured using a miniature gyroscope attached to the chest and a portable recorder placed on the waist. Based on a simple model and the discrete wavelet transform, three parameters related to the PT were measured, namely, the average and standard deviation of transition duration and the occurrence of abnormal successive transitions (number of attempts to have a successful transition). The comparison between two groups of elderly subjects (with high and low fall-risk) showed that the computed parameters were significantly correlated with the falling risk as determined by the record of falls during the previous year, balance and gait disorders (Tinetti score), visual disorders, and cognitive and depressive disorders (p < 0.01). In this study, the wavelet transform has provided a powerful technique for enhancing the pattern of PT, which was mainly concentrated into the frequency range of 0.04-0.68 Hz. The system is especially adapted for long-term ambulatory monitoring of elderly people.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据