期刊
GEOCHEMISTRY GEOPHYSICS GEOSYSTEMS
卷 3, 期 -, 页码 -出版社
AMER GEOPHYSICAL UNION
DOI: 10.1029/2002GC000310
关键词
silicon; silica; isotope; MC-ICP-MS; plasma
[1] Widespread utilization of silicon isotope ratio variations (delta(29)Si and delta(30)Si) as a geochemical and paleoceanographic tracer has been hampered by the hazardous nature of the analysis, which requires the use of a fluorinating gas. Multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS) provides a safer means of silicon isotope analysis. High backgrounds, long rinsing and sample introduction times, and interferences are significant difficulties. It is possible to work around these obstacles, and the measurement of delta(29)Si may be done to a precision averaging +/- 0.1parts per thousand. The delta(30)Si, which cannot be measured directly due to a NO interference at mass 30, may be extrapolated from the theoretically derived and empirically verified relationship, delta(30)Si = 1.93delta(29)Si. Biological materials (sponge spicules and diatoms) measured via MC-ICP-MS yielded delta(30)Si values in the ranges expected from the prior analysis of similar samples by the method of fluorination.