4.5 Article

Measurement of silicon stable isotope natural abundances via multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS)

期刊

出版社

AMER GEOPHYSICAL UNION
DOI: 10.1029/2002GC000310

关键词

silicon; silica; isotope; MC-ICP-MS; plasma

向作者/读者索取更多资源

[1] Widespread utilization of silicon isotope ratio variations (delta(29)Si and delta(30)Si) as a geochemical and paleoceanographic tracer has been hampered by the hazardous nature of the analysis, which requires the use of a fluorinating gas. Multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS) provides a safer means of silicon isotope analysis. High backgrounds, long rinsing and sample introduction times, and interferences are significant difficulties. It is possible to work around these obstacles, and the measurement of delta(29)Si may be done to a precision averaging +/- 0.1parts per thousand. The delta(30)Si, which cannot be measured directly due to a NO interference at mass 30, may be extrapolated from the theoretically derived and empirically verified relationship, delta(30)Si = 1.93delta(29)Si. Biological materials (sponge spicules and diatoms) measured via MC-ICP-MS yielded delta(30)Si values in the ranges expected from the prior analysis of similar samples by the method of fluorination.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据