Intravascular ultrasound findings in the multicenter, randomized, double-blind RAVEL (RAndomized study with the sirolimus-eluting VElocity balloon-expandable stent in the treatment of patients with de novo native coronary artery Lesions) Trial

期刊

CIRCULATION
卷 106, 期 7, 页码 798-803

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1161/01.CIR.0000025585.63486.59

关键词

stents; restenosis; ultrasonics; drugs

向作者/读者索取更多资源

Background-The goal of this intravascular ultrasound investigation was to provide a more detailed morphological analysis of the local biological effects of the implantation of a sirolimus-eluting stent compared with an uncoated stent. Methods and Results-In the RAVEL trial, 238 patients with single de novo lesions were randomized to receive either an 18-min sirolimus-eluting stent (Bx VELOCITY stent, Cordis) or an uncoated stent (Bx VELOCITY stent). In a subset of 95 patients (sirolimus-eluting stent=48, uncoated stent=47), motorized intravascular ultrasound pullback (0.5 mm/s) was performed at a 6-month follow-up. Stent volumes, total vessel volumes, and plaque-behind-stent volumes were comparable. However, the difference in neointimal hyperplasia (2+/-5 versus 37+/-28 MM3) and percent of volume obstruction (1+/-3% versus 29+/-20%) at 6 months between the 2 groups was highly significant (P<0.001), emphasizing the nearly complete abolition of the proliferative process inside the drug-eluting stent. Analysis of the proximal and distal edge volumes showed no significant difference between the 2 groups in external elastic membrane or lumen and plaque volume at the proximal and distal edges. There was also no evidence of intrastent thrombosis or persisting dissection at the stent edges. Although there was a higher incidence of incomplete stent apposition in the sirolimus group compared with the uncoated stent group (P<0.05), it was not associated with any adverse clinical events at 1 year. Conclusions-Sirolimus-eluting stents are effective in preventing neointimal hyperplasia without creating edge effect and without affecting the plaque burden behind the struts.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据