4.7 Article

Aluminium(III)-selective electrode based on a newly synthesized tetradentate Schiff base

期刊

TALANTA
卷 58, 期 2, 页码 397-403

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0039-9140(02)00290-4

关键词

aluminium ion-selective electrode; PVC membranes; bis(5-phenylazo salicylaldehyde) 2,3-naphthalene diimine (5PHAZOSALNPHN)

向作者/读者索取更多资源

A PVC membrane electrode for aluminium ion based on bis(5-phenyl azo salicylaldehyde) 2,3-naphthalene diimine (5PHAZOSALNPHN) as an ion carrier was developed. The electrode exhibits a Nernstian slope of 19.3 +/- 0.8 mV per decade and a linear range of 5.0 x 10(-6)-1.0 X 10(-2) M for AI(NO,),. The limit of detection is 2.5 x 10(-6) M. It has a fast response time of about 10 s and can be used for at least 10 weeks without observing any deviation. The proposed membrane sensor revealed good selectivity for Al3+ over a wide variety of other metal ions and could be used in pH range of 2.9-5.0. It was used as an indicator electrode in potentiometric titration of aluminium ion. (C) 2002 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据