4.7 Article

Evaluation of vapor generation for the determination of nickel by inductively coupled plasma-atomic emission spectrometry

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 374, 期 2, 页码 196-202

出版社

SPRINGER-VERLAG BERLIN
DOI: 10.1007/s00216-002-1372-1

关键词

chemical vapor generation; nickel; inductively coupled plasma atomic emission spectrometry

向作者/读者索取更多资源

Volatile species of Ni were generated by merging acidified aqueous samples and sodium tetrahydroborate(III) in a continuous flow system. The gaseous analyte was subsequently introduced via a stream of At carrier into the inlet tube of the plasma torch. Inductively coupled plasma atomic emission spectrometry (ICP-AES) was used for detection. The operating conditions (chemical and physical parameters) and the concentrations of different acids were evaluated for the efficient generation of Ni vapor. The detection limit (3 sigma(blank)) was 1.8 ng mL(-1). The precision (RSD) of the determination was 4.2% at a level of 500 ng mL(-1) and 7.3% for 20 ng mL(-1) (n=10). The efficiency of the generation process was estimated to be 51%. The possible interfering effect of transition metals (Cd, Co, Cu, Cr, Fe, Mn, Zn), hydride forming elements (As, Ge, Pb, Sb, Se, Sn, Te), and Hg on Ni signal was examined. This study has demonstrated that Ni vapor generation is markedly free of interferences.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据