4.7 Article

Whole-body skeletal imaging in mice utilizing microPET: optimization of reproducibility and applications in animal models of bone disease

出版社

SPRINGER-VERLAG
DOI: 10.1007/s00259-002-0850-1

关键词

MicroPET; MicroCAT; fluoride ion; bone imaging

资金

  1. NCI NIH HHS [P50 CA86306, R0-1 CA82214, R24 CA92865, U01CA84128] Funding Source: Medline

向作者/读者索取更多资源

The aims were to optimize reproducibility and establish [F-18]fluoride ion bone scanning in mice, using a dedicated small animal positron emission tomography (PET) scanner (microPET) and to correlate functional findings with anatomical imaging using computed tomography microCAT). Optimal tracer uptake time for [F-18]fluoride ion was determined by performing dynamic microPET scans. Quantitative reproducibility was measured using region of interest (ROI)-based counts normalized to (a) the injected dose, (b) integral of the heart time-activity curve, or (c) ROI over the whole skeleton. Bone lesions were repetitively imaged. Functional images were correlated with X-ray and microCAT. The plateau of [F-18]fluoride uptake occurs 60 min after injection. The highest reproducibility was achieved by normalizing to an ROI over the whole skeleton, with a mean percent coefficient of variation [(SD/mean)X100] of <15%-20%. Benign and malignant bone lesions were successfully repetitively imaged. Preliminary correlation of microPET with microCAT demonstrated the high sensitivity of microPET and the ability of microCAT to detect small osteolytic lesions. Whole-body [F-18]fluoride ion bone imaging using microPET is reproducible and can be used to serially monitor normal and pathological changes to the mouse skeleton. Morphological imaging with microCAT is useful to display correlative changes in anatomy. Detailed in vivo studies of the murine skeleton in various small animal models of bone diseases should now be possible.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据