Cesium ion selective electrode based on calix[4]crown ether-ester

期刊

TALANTA
卷 58, 期 3, 页码 445-450

出版社

ELSEVIER
DOI: 10.1016/S0039-9140(02)00310-7

关键词

cesium ion; sensor; calix[4]crown ether-ester

向作者/读者索取更多资源

The potentiometric response characteristics of cesium ion selective PVC membrane electrode employing calix[4]crown ether-ester as an ionophore were investigated. The electrode exhibit a good response for cesium ion over wide concentration range of 5.0 x 10(-6)-1.0 X 10(-1) M With a Nernstian slope of 59 mV per decade. The detection limit of electrode is 5.0 x 10(-6) M. The electrode was found to have selectivity for cesium ion over alkali, alkaline and transition metals. The response time of the electrode is less than 20 s and can be used for more than 4 months without observing any divergence in potentiometric response. The electrode response was stable over wide pH range. (C) 2002 Elsevier Science B.V. All rights reserved.