4.6 Article

Performance of a mixing-type CNC for nanoparticles at low-pressure conditions

期刊

JOURNAL OF AEROSOL SCIENCE
卷 33, 期 10, 页码 1389-1404

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0021-8502(02)00092-7

关键词

-

向作者/读者索取更多资源

A mixing-type condensation nucleus counter (MTCNC) was used to detect nanometer-sized particles suspended under low-pressure conditions. The experimental results show that MTCNC is capable of measuring nanoparticles smaller than 10 nm at low pressures above 65 Torr (8.644 kPa) by adjusting the rate of the aerosol flow, the amount of vapor and temperature depending on the desired operation pressure. At pressures above 160 Torr (21.33 kPa), the optimal operation conditions of a MTCNC, variables such as the saturator temperature, condenser temperature and mixing ratio are similar to those used at atmospheric pressure. Since the condensational growth of particles depends greatly upon the residence time, the current MTCNC needs improvement to increase the growth time under even lower pressure conditions. At the optimal operation conditions, the counting efficiencies for nanoparticle size 3 nm particles are 85% at atmospheric pressure and 40% at an operating pressure of 115 Torr (15.33 kPa), respectively. (C) 2002 Elsevier Science Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据