4.7 Article

Confirmation by 16S rRNA PCR of the COBAS AMPLICOR CT/NG test for diagnosis of Neisseria gonorrhoeae infection in a low-prevalence population

期刊

JOURNAL OF CLINICAL MICROBIOLOGY
卷 40, 期 11, 页码 4056-4059

出版社

AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.40.11.4056-4059.2002

关键词

-

向作者/读者索取更多资源

The COBAS AAIPLICOR CT/NG test is widely used for the diagnosis of Neisseria gonorrhoeae infection using genital swabs or urine samples. Although highly specific, cross-reactivity occurs with sonic nonpathogenic strains of Neisseria and Lactobacillus species. In low-prevalence populations, even highly specific assays may require confirmatory testing of positive results. We assessed the positive predictive value (PPV) of this test in a low-prevalence (0.5%) setting. Genital and urine specimens testing positive using the COBAS AMPLICOR NG test were retested using an investigational 16S rRNA PCR assay. Additionally, 737 specimens were tested in parallel by both culture and the above PCR protocol. Of 9,772 specimens tested in-house, 168 were positive by the AMPLICOR test; in addition, 62 AMPLICOR-positive specimens were referred to our laboratory for confirmatory testing, yielding 230 positive specimens. Of these, 72 were confirmed positive by 16S rRNA PCR, yielding a specificity of 98.7% and a PPV of 31.3%. Specificity was similar for all specimen types, whereas PPV varied with prevalence: specimens from males, females, urine specimens, and genital swabs had PPV's of 70.8, 13.3, 51.9, and 20.1%, respectively. The PPV was higher when the initial AMPLICOR optical density (OD) was greater than or equal to3.5 versus initial and repeat OD readings in an equivocal zone of greater than or equal to0.2 to <3.5 (65.1 versus 10.1%; P < 0.001). On repeat testing of specimens with ODs in the equivocal zone, 54 gave ODs of greater than or equal to0.2 and <2.0, 35 gave ODs of greater than or equal to2.0 and <3.5, and 12 gave ODs of; greater than or equal to3.5, with 3.7, 20, and 33.3% confirmed positive, respectively (P = 0.004). Comparing PCR to culture as the gold standard, specificity increased from 96.8 to 99.9% when 16S rRNA PCR was performed on specimens positive by the COBAS AMPLICOR NG test. Confirmatory testing with a more specific method such as 16S rRNA PCR should be considered in low-prevalence populations, especially for specimens with an OD in the equivocal zone.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据